OUR PRODUCTS  Microwave samples digestion system MS-6 |
 |
The system of MS-6 (Microwave samples digestion system) is designed for expansion (ashing) of samples of food products, food raw materials, soil, ecological, materials, biological medium, etc. when analyzing the chemical composition of samples by different techniques: voltammetry, atomic absorption spectroscopy, spectrophotometry, etc. Also, a Microwave samples digestion system MC-6 can be used in the microwave synthesis of nanostructured compounds. |
A patent for the invention № 2165608
«The method of preparing samples for chemical analysis and a device for its implementation»
Sanitary-epidemiological conclusion
78.01.06.344.P.006413.04.09 number of 15.04.2009g.
IS CH 2.2.4/2.1.8.562-96 and
SanPiN 2.1.8/2.2.4.1383-03
Principle of operation is based on the use of microwave energy for rapid volumetric heating of the samples in sealed containers with the addition of oxidants. Thus, the treatment of samples occurs at higher pressures (up to 14 atm.) and temperatures (up to 200 C0). Mineralization process takes place automatically according to a pre-recorded in a control panel standard program algorithm. During the whole process of mineralization in real-time monitor the pressure inside the container.
|
 |
Universal remote control system
A patent for the invention № 2320998
«Microwave system of sample pretreatment for chemical analysis»
(New in the invention is to create a remote device management system based on microwave wireless remote control panel)
The control console is made on the basis of a personal computer (PDA, laptop, desktop computer), working under the operating system Windows (Mobile 5 or higher, XP or higher) and has a module for the wireless Bluetooth. Use of the remote control makes it possible during the process of mineralization will automatically receive a schedule according to the pressure of time, and then stored in the computer memory.
- On the screen in real-time displays relevant information about the status of
- Data is transferred over the radio channel that provides mobility and ease of operator control devices
- Remote control of one or more devices (up to 10 devices.)
- Range up to 10m.
- User-friendly and customizable graphical interface to the data on the screen
- Treatment and storage of results
|
 |
| Quantity of simultaneous pretreatment sample |
up to 12 |
| Massa of sample |
0,25 - 2 g |
| The consumption of reagents per sample not more |
not more 10 ml |
| Duration of mineralization |
15-40 min |
| Limit the pressure in the container |
16 atm |
| Power consumption |
not more 1000 Wt |
Microwave samples digestion system MS-6 includes:
- Microwave samples digestion system with a pressure sensor 1pcs
- Carousel for installing high-pressure containers 1pcs
- Containers of high pressure in the collection 2 pcs
- Wireless remote control with a recorded program pretreatment 1pcs
- Fume sleeve and flange to connect to exhaust ventilation 1 pcs
- Passport 1pcs
- Instruction manuals 1 pcs
- Methodological Support 1pcs